These are medium to large-sized chambers that are custom-built to ensure that they not only fit optimally in your lab’s stability area but also address your large-scale stability testing requirements.
Newtronic Walk-Ins ensure a smooth air-flow pattern to improve the temperature distribution across the chamber. This has also led to huge load reduction on the heating, cooling systems.
This & many other innovations in the construction & design aspects have ensured dramatic reduction in power consumption of our Walk-Ins.
Newtronic has also been regularly exporting the Walk-In Chambers to all the major markets of the world. We ensure our products meet all kinds of country-specific requirements such as voltage specifications, regulatory needs, container shipping, etc.
Offers high-speed data transfer, surpassing serial (RS-485) communication, and adheres to a global standard for communication protocol, enabling data operations from any PC over the LAN without requiring a separate PC interface.
Completely ready-to-use systems, Automatic switch-over upon failure of primary systems to achieve conditions, Switch-over also upon detection of HP/LP & Internal fault, Scheduled & manual change-overs, Recording of all change-overs in the software.
Complete Operations driven through PLC, No Data Loss: Data Storage of upto 1000 readings on the PLC itself, At an interval of 30 mins, data upto 20 days is stored on the PLC itself, Facility to send emails directly from the PLC.
All operations through a single 4" Color touch screen display, PLC IP Address setting Component Working Hours, Standby system change over, Password protection for set values and calibration mode.
Temperature & Humidity Deviations, Door Opening, Water Supply Failure, Component Failure, Mains Failure.
Mobile alert system, Time of alarm transmission, Name of equipment, Transmitted message(Equipment name, alarm time, type of alarm & limits).
IQ, OQ, PQ, Calibration certificates, Operations & maintenance manuals, GA & Electrical drawings, Material test certificates, Training to end-users.
Temperature Overshoot/Undershoot, Humidity Overshoot, Low water-level, Delay Timer to safeguard the Compressor, MCB to cut-off over-current supply, Standby Controlling Sensor upon failure of Primary, In Walk-in chambers door can be open from inside, Emergency Bell switch inside the Chamber.
21 CFR Part 11 Ready, Single software suitable for upto 20 equipment, Automatic daily data back-up, Alarm logs. Alarm acknowledgement with e-signatures, Audit trails, MKT Calculations, Graphical viewing & data trending, Data export in pdf format.
Password based access control & Restricts access to unauthorized people.
When an alarm is triggered, an e-mail is sent to the registered PC or mobile address, *connection to a mail server is required to use e-mail alert.
Temperature Range | 20°C to 60°C |
Temperature Accuracy | ± 0.5°C |
Temperature Uniformity | ± 1°C |
Humidity Range | 40% RH to 95% RH |
Humidity Accuracy | ± 2% RH |
Humidity Uniformity | ± 3% RH |
Chamber Operations based on Ecotronic system incorporating the principles of Dew point control as per the Psychometric Chart | Double walled with 60mm PUF insulated modular panels with easy on-site assembly |
SS304 Interior & Exterior either SS304 or Pre Coated GI | Stainless Steel Trays |
Observation window - 450 x 300 mm | PVC strip curtain for reducing temperature leakages |
Heater line along the door to avoid condensation | U-shaped SS tubular heaters |
Forced air circulation for uniform temperature | Humidity is primarily controlled through the principles of Dew point control (for capacities>8KL) |
CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser | System is made up of split type condenser unit in well-ventilated area to avoid throw of hot air in the room |
Stand by humidity system with booster humidification | Complete Operations driven through PLC with data logging capacity of upto 1000 readings |
High speed Ethernet based communication | All operations through a single 7" Color Touch Screen |
High quality imported make T/Rh combo sensor with 4-20 mA output | Safety devices: Low & High temperature cut-off settable thermostats |
Audio/Visual alarms for deviations | Complete with DQ, IQ, OQ, PQ & SOP & Operational Manual |
Power Supply: 230/415 volts 50 Hz main supply. Also available with 60 Hz Cycle.
ICDAS - A Single Software installation for all Newtronic make equipments | 8 pt. Temperature & Humidity Scanner |
Stand by Cooling System with auto switch over | Door Access system (Numeric or Biometric) |
Remote alarm system suitable for upto 4 nos. Newtronic make equipments | Mobile Alert System |
Model | Capacity | Internal Size | External Size | Trays Qty. | Trays Size | |
---|---|---|---|---|---|---|
Cu.Ft. | Liter | (H*W*D) mm | (H*W*D) mm | (W*D) mm | ||
NLWH208SI | 208 | 5,848 | 2000 X 1360 X 2150 | 2120 X 1480 X 2270 | 20 | 900 X 300 |
NLWH250SI | 250 | 7,017 | 2400 X 1360 X 2150 | 2520 X 1480 X 2270 | 28 | 900 X 300 |
NLWH307SU | 307 | 8,600 | 2000 X 2000 X 2150 | 2120 X 2120 X 2270 | 20 | 900 X 600 |
NLWH368SU1 | 368 | 10,320 | 2400 X 2000 X 2150 | 2520 X 2120 X 2270 | 28 | 900 X 600 |
NLWH543SU | 543 | 15,216 | 2400 X 2000 X 3170 | 2520 X 2120 X 3290 | 42 | 900 X 600 |
NLWH679SU | 679 | 19,020 | 2400 X 2500 X 3170 | 2520 X 2620 X 3290 | 42 | 900 X 800 |
NLWH1092SU1 | 1092 | 30,600 | 2400 X 2500 X 5100 | 2520 X 2620 X 5220 | 70 | 900 X 800 |
NLWH1787SU1 | 1787 | 50,040 | 3000 X 4000 X 4170 | 3120 X 4120 X 4290 | 126 | 900 X 600 |
NLWH2857SU | 2857 | 80,019 | 3000 X 5100 X 5230 | 3120 X 5220 X 5350 | 162 | 900 X 800 |
For all your industrial chamber needs or to explore collaboration opportunities, don’t hesitate to reach out. We’re here to assist you in any way we can.