This unit controls light and temperature conditions through an easy-to-use color touch-screen interface. Uniform light distribution and high intensity levels allow quick response for forced degradation testing and confirmatory studies. The confirmatory studies should identify precautionary measures needed in manufacturing or in formulation of the product, and if light resistant packaging is needed.
At the end of the exposure period, the samples should be examined for any changes in physical properties (e.g., appearance, clarity, or colour of of solution) and for assay degradation.
Newtronic chambers are fully controlled via the PLC with features such as alarms for aging lamps, auto-switch off of UV lights upon door opening. Specular reflective surfaces provide uniform UVA & VIS light reflection and maintain proper SPD (spectral power distribution). Additionally, the chamber also provides precise humidity control.
Offers high-speed data transfer, surpassing serial (RS-485) communication, and adheres to a global standard for communication protocol, enabling data operations from any PC over the LAN without requiring a separate PC interface.
Mobile Alert system, Time of Alarm transmission, Name of Equipment, Transmitted Message (Equipment Name, Alarm time, Type of Alarm & Limits)
21 CFR Part 11 Ready, Single software suitable for upto 20 equipments, Automatic daily data back-up, Alarm logs. Alarm acknowledgement with e-signatures, Audit trails, MKT Calculations, Graphical viewing & data trending, Data export in pdf format.
IQ, OQ, PQ, Calibration certificates, Operations & maintenance manuals, GA & Electrical drawings, Material test certificates, Training to end-users.
Temperature & Humidity Deviations, Door Opening, Water Supply Failure, Component Failure, Mains Failure.
Temperature Overshoot/Undershoot, Humidity Overshoot, Low water-level, Delay Timer to safeguard the Compressor, MCB to cut-off over-current supply, Standby Controlling Sensor upon failure of Primary, In Walk-in chambers door can be open from inside, Emergency Bell switch inside the Chamber.
When an alarm is triggered, an e-mail is sent to the registered PC or mobile address, connection to a mail server is required to use e-mail alert.
Complete Operations driven through PLC, No Data Loss: Data Storage of upto 1000 readings on the PLC itself, At an interval of 30 mins, data upto 20 days is stored on the PLC itself, Facility to send emails directly from the PLC.
All operations through a single 4" Color touch screen display, PLC IP Address setting, Component Working Hours, Standby system change over, Password protection for set values and calibration mode.
Temperature Range | 10°C to 40°C |
Temperature Accuracy | ± 0.5°C |
Temperature Uniformity | ± 1°C |
Double walled with PUF insulation | Stainless Steel trays |
SS 304 Interior & Exterior either SS 304 or Mild Steel. Additionally Interior walls are flushed with mirrors | Full-length inner glass door, metal door with gasket & lock on the outside. |
Forced air circulation for uniform temperature | PU wheels for easy movement |
Fluorescent LED light fitted on the top of the inner chamber to provide illumination & produce visible light measured in lux | UV light of wavelength 365nm (measured in watt/sq. m.) fitted on top of the chamber U-shaped SS tubular heaters. |
CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser. | Complete Operations driven through PLC with data logging capacity of upto 1000 readings |
High speed Ethernet based communication. | All operations through a single 7″ Color Touch Screen. |
High-quality PT-100 Temperature sensor. | Automatic switch off of UV lights upon door opening. |
Safety devices: Low & high temperature cut-off settable thermostats. | Safety thermostat switches off the lights upon temperature overshoot. |
Audio/Visual alarms for deviations. | Complete with DQ, IQ, OQ, PQ & SOP & Maintenance Manual |
ICDAS – A Common Software suitable for 20 nos. Newtronic make equipments | Door Access system (Numeric or Biometric) |
LUX and UV meter – Hand held or Real-time measurement by inbuilt sensor | Remote alarm system suitable for 4 nos. Newtronic make equipments |
Mobile Alert system |
Model | Capacity – CuFt. | Capacity – Litres | Internal Size – (HxWxD) mm | External Size – (HxWxD) mm | Trays Qty. | Tray Size – (WxD) mm | Approx. Weight – Eqip. Kg | Approx. Weight – Ship. Kg |
---|---|---|---|---|---|---|---|---|
NLPS4SI | 4 | 100 | 450 x 450 x 480 | 925 X 700 X 720 | 1 | 410 x 440 | 125 | 220 |
NLPS8SI | 8 | 227 | 600 x 630 x 600 | 1410 X 790 X 1055 | 1 | 580 x 560 | 175 | 250 |
NLPSH8SI | 8 | 227 | 600 x 630 x 600 | 1410 X 790 X 1075 | 1 | 580 x 560 | 185 | 260 |
Temperature Range | 20°C to 60°C |
Temperature Accuracy | ± 0.5°C |
Temperature Uniformity | ± 1°C |
Humidity Range | 40% RH to 95% RH |
Humidity Accuracy | ± 2% RH |
Humidity Uniformity | ± 3% RH |
Double walled with PUF insulation. | Stainless Steel trays |
SS 304 Interior & Exterior either SS 304 or Mild Steel | Full-length inner glass door & metal door with gasket & lock on the outside |
Internal LED lights | Forced air circulation for uniform temperature |
PU wheels for easy movement | U-shaped SS tubular heaters |
Non condensing steam injection type humidity system with standby heater & auto switch overs | CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser |
Stand by humidity system with booster humidification | Complete Operations driven through PLC with data logging capacity of upto 1000 readings |
High speed Ethernet based communication | All operations through a single 7″ Color Touch Screen |
High quality imported make T/Rh combo sensor with 4-20 mA output | Safety devices: Low & high temperature cut-off settable thermostats |
Audio/Visual alarms for deviations | Complete with DQ, IQ, OQ, PQ & SOP & Maintenance Manual |
ICDAS – A Common Software suitable for 20 nos. Newtronic make equipments | Door Access system (Numeric or Biometric) |
4pt. Temperature & Humidity Scanner | Mobile Alert system |
Stand by Cooling System with auto switch overs (430 litre & above) | Remote alarm system suitable for 4 nos. Newtronic make equipments |
Model | Capacity – CuFt. | Capacity – Litres | Internal Size – (HxWxD) mm | External Size – (HxWxD) mm | Trays Qty. | Tray Size – (WxD) mm | Approx. Weight – Eqip. Kg | Approx. Weight – Ship. Kg |
---|---|---|---|---|---|---|---|---|
NLHC7SI | 7 | 200 | 800 x 500 x 500 | 1315 x 660 x 985 | 3 | 430 x 460 | 150 | 225 |
NLHC16SI | 16 | 430 | 1200 x 600 x 600 | 1715 x 760 x 1155 | 4 | 530 x 560 | 250 | 325 |
NLHC26SI | 26 | 735 | 1500 x 700 x 700 | 2070 x 860 x 1255 | 5 | 630 x 660 | 275 | 350 |
NLHC34SI | 34 | 1000 | 1500 x 800 x 800 | 2070 x 960 x 1355 | 5 | 730 x 760 | 350 | 425 |
NLHC54SI | 54 | 1500 | 1500 x 1600 x 600 | 2070 x 1760 x 1155 | 10 | 695 x 560 | 350 | 450 |
NLHC71SI | 71 | 2000 | 1500 x 1600 x 800 | 2070 x 1760 x 1355 | 10 | 695 x 760 | 375 | 475 |