The chamber also provides an environment for the storage and preservation of gametes and animal tissue cell cultures intended for research at body temperature or under hypoxic conditions.
A unique blend of product features to promote cell growth through the precise control of humidity, temperature, and C02 gas while minimizing the potential for contamination creating the most reliable conditions.
Temperature & Humidity Deviations, Door Opening, Water Supply Failure, Component Failure, Mains Failure.
IQ, OQ, PQ, Calibration certificates, Operations & maintenance manuals, GA & Electrical drawings, Material test certificates, Training to end-users.
Temperature Overshoot/Undershoot, Delay Timer to safeguard the Compressor, MCB to cut-off over-current supply, Standby Controlling Sensor upon failure of Primary, Emergency Bell switch inside the Chamber.
Temperature Range | 18°C to 50°C |
CO₂ Range | 0% to 20% |
SS 304 interior and exterior either SS 304 or mild steel | Tightly sealed inner door made of tempered safety glass |
Direct heating via heaters wound around the chamber | Perforated SS 304 shelves |
Humidification system with condensation protection | Auto sterilization with hot air at 180°C along with Air Jacket formations |
Adjustable heat sterilization cycle | Complete prevention of desiccation – no matter where your cultures are |
Fan assisted air circulation to ensure superior uniformity, rapid recovery and minimum variation between cultures | Complete operations are driven through a PLC controlling system with a colour touch screen display |
Superior mixing of CO₂ gas via specially designed air flow patterns | IR-based CO₂ sensor for CO₂ monitoring and controlling (upto 20%) |
Audio visual alarms for deviations |
ICDAS – A single software installation for all Newtronic make equipments | Mobile Alert system |
Models | Capacity – CuFt. | Capacity – Litres | Internal Chamber – (HxWxD) mm | External Chamber – HxWxD (mm) | Qty of Shelf | Shelf Size – (WxD) mm |
---|---|---|---|---|---|---|
NECO3MI | 3 | 100 | 500 X 450 X 450 | 845 X 590 X 765 | 2 | 430 X 430 |
NECO6MI | 6 | 170 | 600 X 600 X 470 | 945 X 740 X 770 | 2 | 580 X 450 |
Temperature Range | 20°C to 60°C |
Temperature Accuracy | ± 0.5°C |
Temperature Uniformity | ± 1°C |
Humidity Range | 40% RH to 95% RH |
Humidity Accuracy | ± 2% RH |
Humidity Uniformity | ± 3% RH |
Double walled with PUF insulation. | Stainless Steel trays |
SS 304 Interior & Exterior either SS 304 or Mild Steel | Full-length inner glass door & metal door with gasket & lock on the outside |
Internal LED lights | Forced air circulation for uniform temperature |
PU wheels for easy movement | U-shaped SS tubular heaters |
Non condensing steam injection type humidity system with standby heater & auto switch overs | CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser |
Stand by humidity system with booster humidification | Complete Operations driven through PLC with data logging capacity of upto 1000 readings |
High speed Ethernet based communication | All operations through a single 7″ Color Touch Screen |
High quality imported make T/Rh combo sensor with 4-20 mA output | Safety devices: Low & high temperature cut-off settable thermostats |
Audio/Visual alarms for deviations | Complete with DQ, IQ, OQ, PQ & SOP & Maintenance Manual |
ICDAS – A Common Software suitable for 20 nos. Newtronic make equipments | Door Access system (Numeric or Biometric) |
4pt. Temperature & Humidity Scanner | Mobile Alert system |
Stand by Cooling System with auto switch overs (430 litre & above) | Remote alarm system suitable for 4 nos. Newtronic make equipments |
Model | Capacity – CuFt. | Capacity – Litres | Internal Size – (HxWxD) mm | External Size – (HxWxD) mm | Trays Qty. | Tray Size – (WxD) mm | Approx. Weight – Eqip. Kg | Approx. Weight – Ship. Kg |
---|---|---|---|---|---|---|---|---|
NLHC7SI | 7 | 200 | 800 x 500 x 500 | 1315 x 660 x 985 | 3 | 430 x 460 | 150 | 225 |
NLHC16SI | 16 | 430 | 1200 x 600 x 600 | 1715 x 760 x 1155 | 4 | 530 x 560 | 250 | 325 |
NLHC26SI | 26 | 735 | 1500 x 700 x 700 | 2070 x 860 x 1255 | 5 | 630 x 660 | 275 | 350 |
NLHC34SI | 34 | 1000 | 1500 x 800 x 800 | 2070 x 960 x 1355 | 5 | 730 x 760 | 350 | 425 |
NLHC54SI | 54 | 1500 | 1500 x 1600 x 600 | 2070 x 1760 x 1155 | 10 | 695 x 560 | 350 | 450 |
NLHC71SI | 71 | 2000 | 1500 x 1600 x 800 | 2070 x 1760 x 1355 | 10 | 695 x 760 | 375 | 475 |