COOLING (LABORATORY REFRIGERATOR) CHAMBERS

Protecting the integrity of your work is imperative which is why laboratories rely only on NEWTRONIC for their expertise in high quality refrigeration solutions. Our full line of refrigerators provide superior temperature uniformity, reliability and an eco-friendly construction for the safe storage of working and reference standards, vaccines, reagents, raw materials and other critical samples.

NEWTRONIC chambers have become the defacto standard for stability studies of Refrigerated Products.

They also come with specially designed trays for segregating Working Standards & Reference Standards samples for effective storage & quick retrieval.

The chambers are connected through Ethernet to state-of-the-art 21 CFR Part 11 ready software that provide tools for viewing, trending, alarm management, audit trails, MKT & many other such features.

Remember, your lab is not a kitchen! And therefore, domestic fridges can never be a substitute to industrial refrigerators.

Download send Inquiry

Features

ETHERNET COMMUNICATION :

  • High-speed data transfer
  • Significantly faster than serial(RS-485) communication
  • Global standard for communication protocol
  • No need for a separate PC interface
  • Data Operations from any PC over the LAN

STANDBY COOLING SYSTEM

  • Completely ready-to-use systems
  • Automatic switch-over upon failure of primary systems to achieve conditions
  • Switch-over also upon detection of HP/LP & Internal fault
  • Scheduled & manual change-overs
  • Recording of all change-overs in the software

TOUCH SCREEN DOOR ACCESS SYSTEM

  • Temperature & Deviations
  • Door Opening
  • Component Failure
  • Mains Failure

ALARM TRANSMISSION

  • Mobile alert system
  • Time of alarm transmission
  • Name of equipment
  • Transmitted message(Equipment name, alarm time,type of alarm & limits)

MOBILE ALERT

  • Time of Alarm transmission
  • Name of Equipment
  • Transmitted message(Equipment name, alarm time,type of alarm & limits)

SAFETY

  • Temperature Overshoot/Undershoot
  • Delay Timer to safeguard the Compressor
  • MCB to cut-off over-current supply
  • Standby Controlling Sensor upon failure of Primary
  • Emergency Bell switch inside the Chamber

ICDAS SOFTWARE

  • 21 CFR Part 11 Ready
  • Single software suitable for upto 20 equipments
  • Automatic daily data back-up
  • Alarm logs. Alarm acknowledgement with e-signatures
  • Audit trails
  • MKT Calculations
  • Graphical viewing & data trending
  • Data export in pdf format

DOCUMENTATION & TRAINING

  • IQ, OQ, PQ
  • Calibration certificates
  • Operations & maintanance manuals
  • GA & Electrical drawings
  • Material test certificates
  • Training to end-users
Temperature Range 2°C to 8°C
Temperature Accuracy ± 0.5°C
Temperature Uniformity ± 1°C
Double walled with 80mm PUF insulated modular panels with easy on-site assembly System is made up of split type condenser unit in well-ventilated area to avoid throw of hot air in the room
SS 304 Interior & Exterior either SS 304 or Powder Coated G.I. Stand by humidity system
Stainless Steel Trays Complete Operations driven through PLC with data logging capacity of upto 1000 readings & email facility
Observation window - 300 x 300 mm High speed Ethernet based communication
Illuminated working area All operations through a single 4" Color Touch Screen
U-shaped SS tubular heaters High-quality German Make temperature humidity sensor with 4-20mA output
Flooring of Linoleum carpet Safety devices: Low & High temperature cut-off settable thermostats & high humidity cut-off device
Forced air circulation for uniform temperature Audio/Visual alarms for deviations
Non Condensing steam injection type humidity system with standby heater & auto switch overs Complete with IQ, OQ, PQ & Operations & Maintenance Manual
CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser

Power supply: 230/415 volts 50 Hz main supply. Also available with 60 Hz Cycle.
ICDAS - A Common Software suitable for 20 nos. Newtronic make equipments Door Access system
4pt. Temperature Scanners Mobile Alert system
Stand by Cooling System with auto switch over (430 litres & above) Remote alarm system suitable for 4 nos. Newtronic make equipments
Grid-type partition for WS/RS sample arrangement
Model Capacity Internal Size External Size Trays Qty. Tray size Approx. Weight (Kg) Approx. Weight (Kg)
CuFt. Liter (H*W*D )mm (H*W*D) mm (W*D)mm Eqip. Ship.
NLCC7SI 7 200 800 x 500 x 500 1295 x 640 x 990 3 430 X 460 150 225
NLCC16SI 16 430 1200 x 600 x 600 1695 x 740 x 1110 4 530 X 560 250 325
NLCC26SI 26 735 1500 x 700 x 700 2050 x 840 x 1210 5 630 X 660 275 350
NLCC34SI 34 1000 1500 x 800 x 800 2050 x 940 x 1310 5 730 X 760 350 425
NLCC54SI 54 1500 1500 x 1600 x 600 2050 x 1740 x 1125 10 695 X 560 350 450