The Ecotronic System – based on the principles of Dew Point Control as per the Psychrometric Chart has resulted in a reduction of Power Consumption by 70% and Water Consumption by 90%.
Over the years, our Investment in R&D has resulted in developing technology aimed at Energy Optimisation.
Offers high-speed data transfer, surpassing serial (RS-485) communication, and adheres to a global standard for communication protocol, enabling data operations from any PC over the LAN without requiring a separate PC interface.
IQ, OQ, PQ, Calibration certificates, Operations & maintenance manuals, GA & Electrical drawings, Material test certificates, Training to end-users.
Mobile alert system, Time of alarm transmission Name of equipment, Transmitted message(Equipment name, alarm time, type of alarm & limits).
Complete Operations driven through PLC, No Data Loss: Data Storage of upto 1000 readings on the PLC itself, At an interval of 30 mins, data upto 20 days is stored on the PLC itself, Facility to send emails directly from the PLC.
Temperature & Humidity Deviations, Door Opening, Water Supply Failure, Component Failure, Mains Failure.
All operations through a single 4" Color touch screen display, PLC IP Address setting, Component Working Hours, Standby system change over, Password protection for set values and calibration mode.
21 CFR Part 11 Ready, Single software suitable for upto 20 equipments, Automatic daily data back-up, Alarm logs. Alarm acknowledgement with e-signatures, Audit trails, MKT Calculations, Graphical viewing & data trending, Data export in pdf format.
Completely ready-to-use systems, Automatic switch-over upon failure of primary systems to achieve conditions, Switch-over also upon detection of HP/LP & Internal fault, Scheduled & manual change-overs, Recording of all change-overs in the software.
When an alarm is triggered, an e-mail is sent to the registered PC or mobile address, Connection to a mail server is required to use e-mail alert.
Temperature Overshoot/Undershoot, Humidity Overshoot, Low water-level, Delay Timer to safeguard the Compressor, MCB to cut-off over-current supply, Standby Controlling Sensor upon failure of Primary, In Walk-in chambers door can be open from inside, Emergency Bell switch inside the Chamber
Password based access control, Restricts access to unauthorized persons.
Temperature Range | 20°C to 60°C |
Temperature Accuracy | ± 0.5°C |
Temperature Uniformity | ± 1°C |
Humidity Range | 40% RH to 95% RH |
Humidity Accuracy | ± 2% RH |
Humidity Uniformity | ± 3% RH |
Chamber Operations based on Ecotronic system incorporating the principles of Dew point control as per the Psychrometric Chart | Humidity is primarily controlled through the principles of Dew point control (for capacities>8KL) |
Double walled with 60mm PUF insulated modular panels with easy on-site assembly | CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser |
SS 304 Interior & Exterior either SS 304 or Pre Coated GI | System is made up of split type condenser unit in well-ventilated area to avoid throw of hot air in the room |
Stainless Steel Trays | Stand by humidity system with booster humidification |
Observation window – 300 x 300 mm | Complete Operations driven through PLC with data logging capacity of upto 1000 readings |
PVC strip curtain for reducing temperature leakages | High speed Ethernet based communication |
Illuminated working area | All operations through a single 7″ Color Touch Screen |
Flooring of Linoleum carpet | High quality imported make T/Rh combo sensor with 4-20 mA output |
Heater line along the door to avoid condensation | Safety devices: Low & High temperature cut-off settable thermostats |
U-shaped SS tubular heaters | Audio/Visual alarms for deviations |
Forced air circulation for uniform temperature | Complete with DQ, IQ, OQ, PQ & SOP & Maintenance Manual |
ICDAS – A Single Software installation for all Newtronic make equipments | Door Access system (Numeric or Biometric) |
8pt. Temperature & Humidity Scanner | Mobile Alert System |
Stand by Cooling System with auto switch over | Remote alarm system suitable for upto 4 nos. Newtronic make equipments |
Model | Capacity – CuFt. | Capacity – Litres | Internal Size – (HxWxD) mm | External Size – (HxWxD) mm | Tray Qty. | Tray Size – (WxD) mm |
---|---|---|---|---|---|---|
NLWH208SI | 208 | 5,848 | 2000 X 1360 X 2150 | 2120 X 1480 X 2270 | 20 | 900 X 300 |
NLWH250SI1 | 250 | 7,017 | 2400 X 1360 X 2150 | 2520 X 1480 X 2270 | 28 | 900 X 300 |
NLWH307SU | 307 | 8,600 | 2000 X 2000 X 2150 | 2120 X 2120 X 2270 | 20 | 900 X 600 |
NLWH368SU1 | 368 | 10,320 | 2400 X 2000 X 2150 | 2520 X 2120 X 2270 | 28 | 900 X 600 |
NLWH543SU | 534 | 15,216 | 2400 X 2000 X 3170 | 2520 X 2120 X 3290 | 42 | 900 X 600 |
NLWH679SU | 679 | 19,020 | 2400 X 2500 X 3170 | 2520 X 2620 X 3290 | 42 | 900 X 800 |
NLWH1092SU1 | 1092 | 30,600 | 2400 X 2500 X 5100 | 2520 X 2620 X 5220 | 70 | 900 X 800 |
NLWH1787SU1 | 1787 | 50,040 | 3000 X 4000 X 4170 | 3120 X 4120 X 4290 | 126 | 900 X 600 |
NLWH2857SU | 2857 | 80,019 | 3000 X 5100 X 5230 | 3120 X 5220 X 5350 | 162 | 900 X 800 |
Temperature Range | 20°C to 60°C |
Temperature Accuracy | ± 0.5°C |
Temperature Uniformity | ± 1°C |
Humidity Range | 40% RH to 95% RH |
Humidity Accuracy | ± 2% RH |
Humidity Uniformity | ± 3% RH |
Double walled with PUF insulation. | Stainless Steel trays |
SS 304 Interior & Exterior either SS 304 or Mild Steel | Full-length inner glass door & metal door with gasket & lock on the outside |
Internal LED lights | Forced air circulation for uniform temperature |
PU wheels for easy movement | U-shaped SS tubular heaters |
Non condensing steam injection type humidity system with standby heater & auto switch overs | CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser |
Stand by humidity system with booster humidification | Complete Operations driven through PLC with data logging capacity of upto 1000 readings |
High speed Ethernet based communication | All operations through a single 7″ Color Touch Screen |
High quality imported make T/Rh combo sensor with 4-20 mA output | Safety devices: Low & high temperature cut-off settable thermostats |
Audio/Visual alarms for deviations | Complete with DQ, IQ, OQ, PQ & SOP & Maintenance Manual |
ICDAS – A Common Software suitable for 20 nos. Newtronic make equipments | Door Access system (Numeric or Biometric) |
4pt. Temperature & Humidity Scanner | Mobile Alert system |
Stand by Cooling System with auto switch overs (430 litre & above) | Remote alarm system suitable for 4 nos. Newtronic make equipments |
Model | Capacity – CuFt. | Capacity – Litres | Internal Size – (HxWxD) mm | External Size – (HxWxD) mm | Trays Qty. | Tray Size – (WxD) mm | Approx. Weight – Eqip. Kg | Approx. Weight – Ship. Kg |
---|---|---|---|---|---|---|---|---|
NLHC7SI | 7 | 200 | 800 x 500 x 500 | 1315 x 660 x 985 | 3 | 430 x 460 | 150 | 225 |
NLHC16SI | 16 | 430 | 1200 x 600 x 600 | 1715 x 760 x 1155 | 4 | 530 x 560 | 250 | 325 |
NLHC26SI | 26 | 735 | 1500 x 700 x 700 | 2070 x 860 x 1255 | 5 | 630 x 660 | 275 | 350 |
NLHC34SI | 34 | 1000 | 1500 x 800 x 800 | 2070 x 960 x 1355 | 5 | 730 x 760 | 350 | 425 |
NLHC54SI | 54 | 1500 | 1500 x 1600 x 600 | 2070 x 1760 x 1155 | 10 | 695 x 560 | 350 | 450 |
NLHC71SI | 71 | 2000 | 1500 x 1600 x 800 | 2070 x 1760 x 1355 | 10 | 695 x 760 | 375 | 475 |